Authors

RSS
Oscillometry

Oscillometry

A month or two ago in the AARC Diagnostics forum several members noted that their labs had acquired Impulse Oscillometry systems a number of years ago but that their physicians had since stopped ordering oscillometry tests, mostly because nobody understood what it was measuring and didn’t know how to interpret the results. There are a number of reasons why this is probably not an uncommon scenario and why, despite being first described in 1956, oscillometry is not used more widely.

13 August 2017 by [Richard Johnson]  •  0 comments

Eye was fooled

Eye was fooled

A couple of days ago I was reviewing (triaging, actually) the spirometry portion of a full panel of PFTs performed with pretty terrible test quality and was trying to decide if the technician responsible for performing the tests had made the right selections from the patient’s test results. I noticed that the FEV1 that had been selected was actually the lowest FEV1 from the all the spirometry efforts the patient made, and was trying to decide whether this was really the correct choice. We use peak flow to help determine which FEV1 to select and that particular spirometry effort appeared to have the highest and sharpest peak flow by a large margin:

29 July 2017 by [Richard Johnson]  •  0 comments

PFTs on YouTube

PFTs on YouTube

A friend recently sent me the links to several YouTube videos on pulmonary function testing. I’ve spent some time off and on over the last year looking at YouTube videos and in particular I’ve been looking for ones that can be used as part of technician education. Maybe I’ve set the bar too high but all too often I’ve been disappointed and frustrated with what I’ve found. One reason for this is that many videos are aimed at other audiences than technicians (i.e. medical students, physicians, patients). Another reason is that too often only simple concepts are presented, often in rote fashion and often without good visual explanations (c’mon, these are videos after all, not podcasts). A final reason is that sometimes they’re outdated, misleading or just plain wrong.

23 July 2017 by [Richard Johnson]  •  0 comments

Z Score to remember is -1.645

Z Score to remember is -1.645

The use of Z scores to report PFT results, both clinically and for research is occurring more and more frequently. Both the Z score and the Lower Limit of Normal (LLN) come from the same roots and in that sense can be said to be saying much the same thing. The difference between the two however, is in the emphasis each places on how results are analyzed. The LLN primarily emphasizes only whether a result is normal or abnormal. The Z score is instead a description of how far a result is from the mean value and therefore emphasizes the probability that a result is normal or abnormal.

16 July 2017 by [Richard Johnson]  •  0 comments

Contraindications

Contraindications

A couple weeks ago I was asked whether it was safe for a patient with an abdominal aortic aneurysm (AAA) to have pulmonary function testing. My first thought was that it was probably unsafe but after a moment or two of thought I realized that I hadn’t reviewed the subject for a long time. When I checked the 2005 ATS/ERS general testing guidelines (there are no contraindications in the 2005 spirometry guidelines) I found that AAA wasn’t mentioned at all. In fact, the only absolute contraindication mentioned was that patients with a recent myocardial infarction (<1 month) should not be tested. Some relative contraindications were mentioned:

09 July 2017 by [Richard Johnson]  •  0 comments

Assessing post-BD improvement in FEV1 and FVC as a percent of the predicted

Assessing post-BD improvement in FEV1 and FVC as a percent of the predicted

The 2005 ATS/ERS standards for assessing post-bronchodilator changes in FVC and FEV1 have been criticized numerous times. A recent article in the May issue of Chest (Quanjer et al) has taken it to task on two specific points: the change in FVC and FEV1 has to be at least 200 ml the change is assessed based on the percent change (≥12%) from the baseline value The article points out that the 200 ml minimum change requires a proportionally larger change for a positive bronchodilator response in the short and the elderly. Additionally, by basing the post-BD change on the baseline value it lowers the threshold (in terms of an absolute change) for a positive bronchodilator response as airway obstruction become more severe. As a way of mitigating these problems the article recommends looking at the post-bronchodilator change as a percent of predicted rather than as a percent of baseline.

02 July 2017 by [Richard Johnson]  •  0 comments

What do you do when the predicted is zero?

What do you do when the predicted is zero?

A very strange spirometry report came across my desk a couple of days ago. Observed: Predicted: %Predicted: FVC: 3.07 0 29767 FEV1: 2.15 0 37586 FEV1/FVC: 70 71 101% My first thought was that some of the demographics information had been entered incorrectly but when I checked the patient’s age, height, gender and race all were present, all were reasonably within the normal range for human beings in general and more importantly, all agreed with what was in the hospital’s database for the patient. I tried changing the patient’s height, age, race and gender to see if it would make a difference and although this made small changes in the percent predicted when I did this the predicteds were still zero.

27 June 2017 by [Richard Johnson]  •  0 comments

The effects of anemia on exercise

The effects of anemia on exercise

Last week I was reviewing the exercise test results from a patient that appeared to have a relatively straightforward cardiovascular limitation when I noticed the patient also had severe anemia (Hgb = 7.1). Once that fact came up it was no longer clear the patient actually had a cardiac limitation at all. First the results: Rest: %Predicted: AT: %Predicted: Max: %Predicted: VO2 (LPM): 0.33 13% 0.73 28% 1.45 56% VO2 (ml/kg/min): 5.0 11.0 21.6 VCO2 (LPM) 0.26 0.63 1.81 RER: 0.73 0.83 1.24 SaO2: 98% 97% 97% PetCO2: 35.2 38.6 31.8 Ve/VO2: 34 26 43 Ve/VCO2: 47 31 35 Ve (LPM): 11.6 8% 19.2 13% 62.9 44% Vt (L): 0.78 1.29 2.19 RR: 15 15 29 HR (BPM): 61 35% 92 52% 152 85% BP (mmHg): 92/62 102/64 O2 Pulse (ml/beat): 5.8 39% 8.2 55% 9.8 66% Other values calculated from the data include:

22 May 2017 by [Richard Johnson]  •  0 comments

Is there such a thing as a normal decrease when the FEV1 isn’t normal?

Is there such a thing as a normal decrease when the FEV1 isn’t normal?

I’ve mentioned before that my lab’s database goes back to 1990, so we now have 27 years of test results available for trending. At least a couple times a week we have a patient who was last seen 10 or even 20 years ago. When I review their results I try to see if there has been any significant change from their last tests. Since the last tests are often quite some time in the past the changes in an absolute sense are often noticeably large. The question then becomes whether or not these changes are normal.

14 May 2017 by [Richard Johnson]  •  0 comments

What does an inverse I:E Ratio during exercise mean?

What does an inverse I:E Ratio during exercise mean?

Inspiration and expiration usually take different lengths of time, with inspiration almost always being shorter than exhalation. This is due to both to the physiology of breathing and to the pathophysiology of disease processes. During incremental exercise testing there are usually patterns to the way that inspiratory and expiratory times change and these are occasionally diagnostic. When I started in this field the relationship between inspiratory and expiratory time was usually expressed as the I:E ratio, which was most often written as something like 1:1.2. One of my medical directors pointed out to me that when talking about I:E ratio it was difficult to determine what you meant if you said it was increasing or decreasing. For this reason I started reporting the I:E ratio as the E/I ratio so that instead of 1:1.2 it’s just 1.2.

07 May 2017 by [Richard Johnson]  •  0 comments